Headspace MAP Gas Analyzers

Features

Relevant link:link

Versatile oxygen analyzer system for O2 measurement and shelf life determination.

Benefits

  • Instant results
  • Shelf life testing – continuous O2 monitoring over time
  • Perfect for testing packages with limited headspace
  • Non-invasive measurement

Features

  • Lightweight
  • Non-invasive and optical measurement
  • Invasive testing with needle
  • No gas extraction
  • Fast reading of O2
  • Conforms to ASTM F2714-08
  • Conforms to ASTM F3136-15
Specifications
Detector and base
Warm-up time 20 minutes
Detector dimensions 4.83 x 3.30 x 22.86 cm with needle (HxWxD)
4.83 x 3.30 x 15.24 cm without needle (HxWxD)
Measurement method Fluorescent decay
Power Standard power USB port (2.5 watt)
Operating temperature 10-35°C
Compliances CE/CSA/UL
Pressure compensation needle sensors
Repeatability
(0 to 150 mmHg)
(150 to 1000 mmHg)
± 0.100% (1000 ppm) O2 or 5% of reading, whichever is greater
± 0.015% (150 ppm) O2 or 2% of reading, whichever is greater
Range 0.050% (500 ppm) to 25%
Operating temperature 5-40°C
Sensors:
Adhesive and ImPULSE
Repeatability (certified) ± 0.03% (300 ppm) O2 or 3% of reading, whichever is greater
Range of Adhesive sensor 0.001% (10 ppm) to 25% O2 Permeation Mode
0.03% (300 ppm) 25% Headspace Mode
Operating temperature 5-40°C
Standards ASTM F2714-08, ASTM F3136-15
Standards
  • ASTM F2714-08
  • ASTM F3136-15
Related Products

Advantage

35+ Yrs Integrated Testing Equipment Service

Comprehensive one-stop vertically integrated service to meet the needs of clients at every stage

Exclusive Distributor for International Testing Brands

Exclusive distributor for multiple globally renowned brands, providing clients with the latest testing equipments

Professional After-Sales Support Services

Our engineers possess extensive expertise and experience, delivering precise and efficient technical support