PRODUCTS

  • Product Brand

  • Product Category

  • Industry

OpTech®

Headspace MAP Gas Analyzers

Features

Relevant link:link

Versatile oxygen analyzer system for O2 measurement and shelf life determination.

Benefits

  • Instant results
  • Shelf life testing – continuous O2 monitoring over time
  • Perfect for testing packages with limited headspace
  • Non-invasive measurement

Features

  • Lightweight
  • Non-invasive and optical measurement
  • Invasive testing with needle
  • No gas extraction
  • Fast reading of O2
  • Conforms to ASTM F2714-08
  • Conforms to ASTM F3136-15
Specifications
Detector and base
Warm-up time 20 minutes
Detector dimensions 4.83 x 3.30 x 22.86 cm with needle (HxWxD)
4.83 x 3.30 x 15.24 cm without needle (HxWxD)
Measurement method Fluorescent decay
Power Standard power USB port (2.5 watt)
Operating temperature 10-35°C
Compliances CE/CSA/UL
Pressure compensation needle sensors
Repeatability
(0 to 150 mmHg)
(150 to 1000 mmHg)
± 0.100% (1000 ppm) O2 or 5% of reading, whichever is greater
± 0.015% (150 ppm) O2 or 2% of reading, whichever is greater
Range 0.050% (500 ppm) to 25%
Operating temperature 5-40°C
Sensors:
Adhesive and ImPULSE
Repeatability (certified) ± 0.03% (300 ppm) O2 or 3% of reading, whichever is greater
Range of Adhesive sensor 0.001% (10 ppm) to 25% O2 Permeation Mode
0.03% (300 ppm) 25% Headspace Mode
Operating temperature 5-40°C
Standards ASTM F2714-08, ASTM F3136-15
Standards
  • ASTM F2714-08
  • ASTM F3136-15
Related Products

Advantage

35年+ 檢測設備垂直整合服務

垂直整合的一條龍服務滿足各階段客戶的需求

國際檢測品牌總代理

多個國際知名檢測品牌總代理,不斷為客戶提供最新檢測設備

專業售後支援服務

工程師具有專業知識與經驗,能準確與有效提供技術支援等服務